Validation and comparative analysis of a contemporary non-contact corneal aesthesiometer
Loading...
Author (Corporation)
Publication date
20.04.2026
Type of student thesis
Course of study
Collections
Type
01A - Journal article
Editors
Editor (Corporation)
Supervisor
Parent work
Journal of Clinical Medicine
Special issue
Diagnosis and Treatment of Ocular Surface Diseases
DOI of the original publication
Link
Related research data
Series
Series number
Volume
15
Issue / Number
8
Pages / Duration
Patent number
Publisher / Publishing institution
MDPI
Place of publication / Event location
Edition
Version
Programming language
Assignee
Practice partner / Client
Abstract
Background: Corneal sensitivity is a key indicator of ocular surface health. This prospective, cross-sectional study evaluated agreement between corneal sensitivity thresholds obtained from equivalent stimulus settings on a contemporary, enhanced dual-temperature non-contact corneal aesthesiometer (NCCA) and a previously validated (standard) device.
Methods: Central corneal sensitivity thresholds were measured in the right eyes of healthy participants using both devices. Participants with previous ocular surgery, laser treatment, trauma, contact lens wear, diabetes, or peripheral neuropathy were excluded. Sensitivity thresholds were determined using a forced-response, double-staircase protocol. Inter-device agreement was assessed using Bland–Altman analysis, and consistency was assessed using intraclass correlation coefficients.
Results: Median corneal sensitivity thresholds in 51 healthy participants (32 female, 19 male; mean age: 33 ± 14 years) did not differ between enhanced (0.23 [0.18 to 0.38]) and standard (0.25 [0.15 to 0.35]) NCCA instruments (p = 0.73). Bland–Altman analysis demonstrated moderate inter-device agreement, with a mean difference of −0.01 mbar (95% limits of agreement: −0.41 to 0.39 mbar). Linear regression analysis identified greater measurement discrepancies at higher thresholds (p < 0.05), indicating greater variability in individuals with reduced corneal sensitivity.
Conclusions: The enhanced NCCA yields reliable corneal sensitivity measures for a room-temperature stimulus and acceptable agreement with the existing (standard) model.
Keywords
Subject (DDC)
Event
Exhibition start date
Exhibition end date
Conference start date
Conference end date
Date of the last check
ISBN
ISSN
2077-0383
Language
English
Created during FHNW affiliation
Yes
Strategic action fields FHNW
Publication status
Published
Review
peer-reviewed
Open access category
Gold
Citation
Xue, A. L., Britten-Jones, A. C., Zhuang, D., Jennings, C. J., Müntz, A., Misra, S. L., Downie, L. E., & Craig, J. P. (2026). Validation and comparative analysis of a contemporary non-contact corneal aesthesiometer. Journal of Clinical Medicine, 15(8). https://doi.org/10.3390/jcm15083145