Ultrasound melted polymer sleeve for improved screw anchorage in trabecular bone-A novel screw augmentation technique

dc.accessRightsAnonymous
dc.audienceScience
dc.contributor.authorSchmoelz, Werner
dc.contributor.authorMayr, Raul
dc.contributor.authorSchlottig, Falko
dc.contributor.authorIvanovic, Nikola
dc.contributor.authorHörmann, Romed
dc.contributor.authorGoldhahn, Jörg
dc.date.accessioned2016-03-18T11:11:01Z
dc.date.available2016-03-18T11:11:01Z
dc.date.issued2016-02-23
dc.description.abstractBACKGROUND: Screw anchorage in osteoporotic bone is still limited and makes treatment of osteoporotic fractures challenging for surgeons. Conventional screws fail in poor bone quality due to loosening at the screw-bone interface. A new technology should help to improve this interface. In a novel constant amelioration process technique, a polymer sleeve is melted by ultrasound in the predrilled screw hole prior to screw insertion. The purpose of this study was to investigate in vitro the effect of the constant amelioration process platform technology on primary screw anchorage. METHODS: Fresh frozen femoral heads (n=6) and vertebrae (n=6) were used to measure the maximum screw insertion torque of reference and constant amelioration process augmented screws. Specimens were cut in cranio-caudal direction, and the screws (reference and constant amelioration process) were implanted in predrilled holes in the trabecular structure on both sides of the cross section. This allowed the pairwise comparison of insertion torque for constant amelioration process and reference screws (femoral heads n=18, vertebrae n=12). Prior to screw insertion, a micro-CT scan was made to ensure comparable bone quality at the screw placement location. FINDINGS: The mean insertion torque for the constant amelioration process augmented screws in both, the femoral heads (44.2Ncm, SD 14.7) and the vertebral bodies (13.5Ncm, SD 6.3) was significantly higher than for the reference screws of the femoral heads (31.7Ncm, SD 9.6, p<0.001) and the vertebral bodies (7.1Ncm, SD 4.5, p<0.001). INTERPRETATION: The interconnection of the melted polymer sleeve with the surrounding trabecular bone in the constant amelioration process technique resulted in a higher screw insertion torque and can improve screw anchorage in osteoporotic trabecular bone.
dc.description.urihttp://www.sciencedirect.com/science/article/pii/S0268003316000450
dc.identifier.doi10.1016/j.clinbiomech.2016.02.010
dc.identifier.issn0268-0033
dc.identifier.issn1879-1271
dc.identifier.urihttp://hdl.handle.net/11654/17635
dc.language.isoen
dc.publisherElsevieren_US
dc.relation.ispartofClinical Biomechanicsen_US
dc.subjectBone ameliorationen_US
dc.subjectScrew anchorageen_US
dc.subjectAugmentationen_US
dc.subjectTrabecular boneen_US
dc.subjectInsertion torqueen_US
dc.subjectPrimary stabilityen_US
dc.titleUltrasound melted polymer sleeve for improved screw anchorage in trabecular bone-A novel screw augmentation technique
dc.type01A - Beitrag in wissenschaftlicher Zeitschrift
dc.volume33
dspace.entity.typePublication
fhnw.InventedHereYes
fhnw.IsStudentsWorkno
fhnw.PublishedSwitzerlandNo
fhnw.ReviewTypeAnonymous ex ante peer review of a complete publication
fhnw.affiliation.hochschuleHochschule für Life Sciencesde_CH
fhnw.affiliation.institutInstitut für Chemie und Bioanalytikde_CH
fhnw.pagination79-83
fhnw.publicationOnlineJa
fhnw.publicationStatePublished
relation.isAuthorOfPublicationc9a2b9dd-53af-4d81-b725-3d252f03e94f
relation.isAuthorOfPublication.latestForDiscoveryc9a2b9dd-53af-4d81-b725-3d252f03e94f
Dateien
Lizenzbündel
Gerade angezeigt 1 - 1 von 1
Lade...
Vorschaubild
Name:
license.txt
Größe:
2.94 KB
Format:
Item-specific license agreed upon to submission
Beschreibung: