Simple and rapid high performance liquid chromatography method for the determination of polidocanol as bulk product and in pharmaceutical polymer matrices using charged aerosol detection
Type
01 - Zeitschriftenartikel, Journalartikel oder Magazin
Zusammenfassung
Currently, neither the European nor the United States Pharmacopoeia provide a method for the determination of polidocanol (PD) content despite the fact that PD, besides being an excipient, is also used as an active pharmaceutical ingredient. We therefore developed a method where the PD content was determined using a Kinetex C18 column operated at 40°C with water-acetonitrile (15:85, v/v) as mobile phase. A Corona(®) charged aerosol detector was employed for the detection of PD that is lacking a suitable UV chromophore. The method was fully validated. Additionally, the method was applied for the determination of PD release from a pharmaceutical polymer matrix consisting of poly-ɛ-caprolactone and poly(lactic-co-glycolic acid) and PD.
DOI der Originalausgabe
https://doi.org/10.1016/j.jpba.2014.11.012Übergeordnetes Werk
Journal of Pharmaceutical and Biomedical Analysis
Jahrgang
104
Seiten
17-20
Verlag / Hrsg. Institution
Elsevier