Millirobot magnetic manipulation for ocular drug delivery with sub millimeter precision
Type
04B - Beitrag Konferenzschrift
Zusammenfassung
Significant progress has been made in the development of magnetic micromanipulation for minimally invasive surgery. The development of systems to localize millimetric size robots during magnetic navigation and without line of sight remains however a challenging task. In this study, we focused on the development of a tracking system aiming to fill this gap. A robot which consists of a cylindrical magnet of 1 mm diameter is localized using a 2D array of 3D magneto resistive sensors. The system provides a tracking of the robot with a refreshing rate of 2 Hz. The developed tracking algorithm reaches a mean absolute error for the position and the orientation of, respectively 0.56 mm and 5.13° in 2D. This system can be added to existing magnetic navigation systems allowing closed loop control of the navigation. The presented tracking system makes it possible to target applications such as minimally invasive ocular drug delivery.
DOI der Originalausgabe
https://doi.org/10.1109/SENSORS52175.2022.9967141Übergeordnetes Werk
2022 IEEE Sensors
Verlag / Hrsg. Institution
IEEE
Verlagsort / Veranstaltungsort
Dallas
Veranstaltung
IEEE Sensors;
2022-10-30;
2022-11-02
2022-10-30;
2022-11-02
Herausgeber/Herausgeberin Körperschaft
IEEE